SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001) 43
○ If the frequency reading is reasonable, but the corresponding pressure is not, this indi-
cates that the appropriate calibration is not loaded into the processor. Use the
LDDRUCK
program (see ADVField Software Manual) to load the
DRUCK
calibration.
Show DRUCK or Show PAROSFREQ
• Displays
DRUCK
/
PAROSFREQ
calibration data in the following format:
Druck pressure sensor calibration data
--------------------------------------
SN: 1276950
F0: 1.200e+07
X: 8.4410000e+03
Y: 5.6000000e+02
T0: -1.0000000e+01
D0: 6.2900000e+02
DT: -2.2100000e+00
K00: 1.5962190e+03 K01: -2.6693000e-02
K10: 1.6820332e+00 K11: -1.0395800e-05
K20: 3.9933779e-04 K21: -4.9470601e-09
K30: 8.0327140e-08 K31: -1.4093000e-12
K40: 1.7388157e-11 K41: 2.2691999e-16
K50: 2.5953699e-15 K51: 1.3877999e-19
K02: 3.1005999e-05 K03: 2.6474001e-07
K12: 1.8332999e-08 K13: 8.2459997e-11
K22: 1.6572001e-11 K23: -1.7079999e-14
K32: -7.7020003e-15 K33: -9.9870002e-17
K42: -4.0642002e-18 K43: -1.0678000e-19
K52: 5.3660002e-21 K53: -2.8249999e-23
3.9.5. External Sensor Commands – LISST-100 Sensor
These commands apply to systems that have an external LISST sensor installed (§4.12). The
LISST-100 is a Laser Interferometry and Transmissometry particle distribution measurement de-
vice designed by Sequoia Scientific.
LisstSensorInstalled Set [YES|NO]
• Selects or disables the LISST sensor (§4.12).
• Set at the factory. Does not need to be changed except in some special cases.
• The LISST sensor is a serial instrument that uses the second COM port on the ADVField’s
CPU board. As such, other serial devices (e.g., Paros, MicroCat CTD) must be disabled.
• When set to YES, LisstBufferSize is automatically set to 80 bytes (default).
LisstBufferSize or LBS [d]
• Default parameter:
80
bytes
• Without parameter: returns its current setting.
• When used with a valid parameter, sets the size of the LISST sample buffer.