EasyManua.ls Logo

Zeiss Axio Vert.A1 - 3.1.12 Adjusting the HAL 100 Illuminator

Zeiss Axio Vert.A1
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PUTTING THE INSTRUMENT INTO OPERATION
Carl Zeiss Mounting Standard Components Axio Vert.A1
58 431030-7044-001 05/2012
3.1.12 Adjusting the HAL 100 Illuminator
Coarse Adjustment
Loosen the clamping screw (Fig. 3-21/2) and
then detach the operational halogen illuminator
(Fig. 3-22/4) from the microscope stand.
Switch on the microscope. (See Section 1.1.)
Aim the light beam at a projection surface (wall)
that is at least 3 m away.
In so doing, be sure not to look into
the light exit of the illuminator.
Use ball-headed screwdriver SW 3 to set the
adjusting screw (Fig. 3-22/1) until both lamp
filament images are shown on the projection
surface as sharply defined as possible.
Use the adjusting screws (Fig. 3-22/2 and 3) to
align the lamp filaments and their mirror image
in such a way to each other that the mirror
image fills up the spaces of the lamp filaments
(Fig. 3-22/5).
Fine Adjustment
Reattach the microscope lamp to the microscope stand and fix it into position, using the clamping
screw.
Switch off the diffusion disk (Fig. 2-9/26) and remove the filter from the light path.
Use the 20x objective to focus on a mirror that has been put on and then go to a clean area of the
specimen.
Remove the eyepiece. Use the adjusting screws (Fig. 3-22/2 and 3) to center the lamp filaments and
their mirror image in the pupil image.
Optimize the overlapping, using the adjusting screw (Fig. 3-22/1).
Switch on the diffusion disk and reactivate the filter.
Fig. 3-22 Adjusting the HAL 100 Illuminator

Table of Contents

Related product manuals