9 Technical Data and Conformity | 9.5 Performance Data and Specifications | EVO 25 ZEISS
Parameter Value
A grounding wire AWG10 (≥ 5m) is delivered
with the microscope. It serves as connection
between the grounding screw terminal (Ø
8mm) and the microscope (Ø 6mm).
Cross section: Min. AWG10 (between
grounding screw terminal and PE of the CEE
FEMALE RECEPTACLE)
Circuit breaker type (at house installation) 16A Type C
Protection class Class I
Momentary interruption Less than a half cycle
Ampere interrupting capacity (AIC) Min. 10,000A
rms
Location
Requirements
Parameter Requirement
Installation site Exclusively inside buildings
Recommended
room size
Min. 3.6m × 4.0m × 2.3m
Service area Min. 0.8m at each side
Entrance Min. 0.8m wide
Hallways Min. 1.0m wide
Corners Min. 1.2m wide
Transport ways Free of staircases
Installation cate-
gory
II
Floor stability > 1000kg/m²
Exhaust Line
If toxic chemicals or biological specimens are used an exhaust line is recommended to remove the
waste gas of the pre-vacuum pump and to transmit it to the outside.
Nitrogen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. The nitro-
gen can be taken either from a gas cylinder or from an in-house supply system.
166 Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006