EasyManua.ls Logo

Zeiss EVO - Page 96

Zeiss EVO
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 First Operating Steps | 5.7 Working with Specific Specimen Types ZEISS
15. Adjust the working parameters, refer to Non-conductive Specimen [
}
93].
16. To return to HV mode, repeat steps 1 to 5 and step 7 to dismount the aperture, deselect
the aperture in the Apertures tab and click Go To HV in the Vacuum tab.
5.7.2.2 Aligning the Gun and Mid-column Aperture for EasyVP Mode
To achieve an optimal alignment of the beam through the 20 μm mid-column aperture, you need
to perform an alignment procedure.
Info
You can refer to Using the Automatic Gun Alignment Functions [
}
79] and Using the Auto
Aperture Alignment Function [
}
80] to help you with these tasks.
Info
Once the alignment is optimized, do not change the position of the X and Y micrometer
gauges on the mid-column aperture changer.
Prerequisite
ü
The EasyVP aperture is fitted.
ü
The 20µm aperture is selected at the mid-column aperture changer (position 2).
Procedure
1. From the Menu Bar, select Tools > User Preferences.
à The User Preferences dialog is displayed.
2. Check that Auto Calibration and User Align are set to Yes.
3. Adjust the following settings in the SEM Controls panel:
Select the EasyVP Aperture
Optibeam = Resolution
Aperture Size = 20.00µm
EHT Target = 20.00 kV
I Probe = 100 pA
WD = 8.5 mm
Signal A = SE1
Operating Mode = Normal
INFO: For details on where to adjust these settings, refer to Acquiring an Image [
}
73].
4. To carry out a C1/C2/C3 lens hysteresis removal, press <Shift + F2>.
5. To create an emission image, select Beam > Gun Align from the Menu Bar.
à
The Gun Alignment window is displayed.
96 Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006

Table of Contents

Related product manuals