5 Calibration, M2000
5.2.5 Initialization of Levelmeter 2000
3HAC021111-001 Revision: B 109
Measuring units
Installation of
sensor
Result The Levelmeter 2000 is now initialized and ready for service.
4. If type 5 does not flash, press ZERO/SELECT to
select filter type 5 and press ENTER .
Step Action Info/Illustration
Step Action Info/Illustration
1. Press ON/MODE until the dot flashes under UNIT . Shown in the figure Level-
meter 2000 on page 108!
2. Press ENTER.
3. Press ZERO/SELECT until mm/m flashes.
Two decimals (0.00) are shown on the display
4. Press ENTER.
Step Action Info/Illustration
1. Connect the sensor to the Sensor connection point. Shown in the figure Level-
meter 2000 on page 108!
2. Press ON/MODE .
3. Press ON/MODE until the dot flashes under SEN-
SOR .
4. Press ENTER.
5. Press ZERO/SELECT until a flashing "A" is shown.
6. Press ENTER.
Wait until the "A" flashes again.
7. Press ENTER.