MODULE 5: MISCELLANEOUS
ARCHITECT
®
RSH System Service Training Guide 5-17
201532-102 FOR INTERNAL USE ONLY
Install Sample Handler RSH Covers
Install left side cover
Install right side cover
Install left and right side cover end caps
Complete Calibrations
M&D 1119 Transport Calibration
M&D 3222 RSH Bar Code Calibration
M&D 3317 RSH Test
- Choose one Priority Section and two Routine Bay Positions
M&D 1111 Sample Pipettor Calibration (i2000
SR
®
System)
M&D 1117 STAT Pipettor Calibration (i2000
SR
System Only)
Verify Routine Bay Partition Alignment Procedure P-01
END OF INSTALLATION & PM ACTIVITY