71
2. Create the task
Create the corresponding DUT tasks as described above, and fill out the necessary information, as shown below.
Table 21 Pressure gauge task setting
Subject Valid value Description
Name Numbers, letters, symbols Task name
Task No. Numbers, letters, symbols Task number
Model Numbers, letters, symbols Model of DUT
GP / AP / DP Pressure type of DUT
Input range Numbers Input range of DUT
Accuracy
0.06%,0.1%,0.16%,0.25%,0.4%,
0.6%,1%,1.6%,2.5%,4%,custom
Accuracy of DUT. If choosing custom, the input number means the accuracy
level of the dial pressure gauge. For example: input 1.6 for gauge level of 1.6,
the range is (0.001~100)%
Resolution 0.05/ custom
The minimum resolution of DUT
(custom 0.001~100)
Owner Numbers, letters, symbols Owner of DUT
Location Numbers, letters, symbols Location of DUT
Note Numbers, letters, symbols Additional information