Advanced Automated Osmometer User’s Guide
• Remove the syringe and tubing from the underside of the waste
bottle cap.
• Replace the cap on the waste bottle and tighten the cap securely.
• Prime the system a minimum of 2 cycles using A
2
O System Fluid
(part no. 200222).
5. Clear the Database
The instrument database size should be checked on a weekly basis
and cleared if it exceeds 1,000 records.
• To clear the database, go into the System Configuration and click
on the Clear Database button (Figure 47).
• The next screen (Figure 48) will allow you to export the results
before deleting them.
Figure 47: Clear Database Button