EasyManua.ls Logo

Advanced Instruments A2O - Page 86

Advanced Instruments A2O
157 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model A
2
O
®
Advanced Automated Osmometer Users Guide
Figure 27
: Calibration Log
the measurement of the corresponding replicate is good. An “O” inside
a red square means there was an error. If more than two replicates for a
given calibrator incur an error, you will need to repeat the calibration.
Stop button
The Stop button stops a calibration procedure. No results from the
attempted calibration are used toward the recalibration. The previous
calibration is still used for future measurements.
Calib. Log
Brings up a screen as shown in Figure 27. The screen displays the
results of previous calibrations. The fields in the spreadsheet are
Date/Time, Type, Outcome, Error, User ID, and Duration. The
Export button saves the log in Comma Separated Values (CSV) format
to a USB memory stick if inserted, or to the instrument’s internal mem-
ory.
Assistance
Brings up the Assistance screen (Figure 37), as described later in this
chapter.
64

Table of Contents

Related product manuals