EasyManuals Logo

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 User Manual

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000
162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
Chapter 4 Maintenance Procedures110
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
CAUTION
!
Make sure the filter end of the tubing does not come into contact with anything that might
contaminate it.
16. Drain the water from the supply tank:
a. Tap SUPPLY TANK Drain on the Maintenance screen.
An on-screen message appears (Figure 4.31).
b. Tap OK in the on-screen message.
An on-screen message appears during the draining process (Figure 4.32).
The water in the supply tank is flushed through the wash system and empties into
the waste tank. Wait for the supply tank to drain.
17. Fill the supply tank again with Type 1, >18.2 M
Ω water following the procedure on
page 99. After the supply tank is filled, remove the supply tubing from the supply port.
Figure 4.31 Draining the Supply Tank
Figure 4.32 On-screen Message – Supply Tank Draining

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 and is the answer not in the manual?

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 Specifications

General IconGeneral
BrandAgena BIOSCIENCE
ModelMassARRAY Nanodispenser RS1000
CategoryMeasuring Instruments
LanguageEnglish