Chapter 4 Maintenance Procedures 109
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
Leave the supply tubing attached to the supply port. The supply tank will be filled
again later in this procedure.
CAUTION
!
Make sure that the filter end of the tubing does not come into contact with anything that might
contaminate it.
13. Drain the 65% ethanol solution from the supply tank:
a. Tap SUPPLY TANK Drain on the Maintenance screen.
An on-screen message appears (Figure 4.29).
b. Tap OK in the on-screen message.
An on-screen message appears during the draining process (Figure 4.30).
The 65% ethanol solution flushes through the wash system and empties into the
waste tank. Wait for the supply tank to drain.
14. Drain the waste tank again following the instructions on page 102. After the waste
tank is drained, remove the tubing attached to the waste port.
15. Fill the supply tank with Type 1, >18.2 M
Ω water following the instructions on page 99.
After the tank is filled, leave the supply tubing attached to the supply port. You will fill
the supply tank again later in this procedure.
Figure 4.29 Draining the Supply Tank
Figure 4.30 On-screen Message Supply Tank Draining

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 and is the answer not in the manual?

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 Specifications

General IconGeneral
BrandAgena BIOSCIENCE
ModelMassARRAY Nanodispenser RS1000
CategoryMeasuring Instruments
LanguageEnglish