Chapter 2 Dispensing to SpectroCHIP Arrays42
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
5. Enter the aspirate offset:
a. Tap the aspirate offset value (Figure 2.33).
b. In the number pad that appears, enter the aspirate offset value you calculated
above and tap enter.
6. Return to the Transfer screen:
a. Tap the BACK button on the Method screen (Figure 2.40).
b. Tap Apply, SAVE, or Discard in the warning message that appears.
Apply This option is recommended. It loads the settings to instrument memory
for use in dispensing runs, but does not save them to a method file.
SAVE Saves the settings to a method file and loads them to instrument memory
for use in dispensing runs. See page 72 for more information.
Discard – Discards any changes that were made.
NOTENOTE
If the currently loaded method is a predefined method, you cannot save changes to it. Predefined
method files are read-only. Tapping SAVE in this case does not affect the method file, but will load
the changes into instrument memory (same as Apply).
Figure 2.33 Enter the Aspirate Offset
Figure 2.34 Method Screen, aspirate/dispense tab
Note: This warning message appears
whenever you leave the Method
screen, even if no changes were made.
In that case, tap Discard.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 and is the answer not in the manual?

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 Specifications

General IconGeneral
BrandAgena BIOSCIENCE
ModelMassARRAY Nanodispenser RS1000
CategoryMeasuring Instruments
LanguageEnglish