Chapter 2 Dispensing to SpectroCHIP Arrays48
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
3. The rinse station preparation screen appears (Figure 2.42). Tap OK and observe the
rinse station as it briefly runs (Figure 2.43).
If water flows out of the rinse station chimneys, the rinse station is operating correctly.
4. In the prompt that appears (Figure 2.44):
Tap YES if the rinse station operated properly.
Tap RETRY to repeat the rinse station test.
Tap NO if the rinse station did not operate properly. See Appendix D, page 151
Figure 2.42 Rinse Station Preparation Screen-1
Figure 2.43 Nanodispenser RS1000 Rinse Station
Rinse station