Agilent 1200 Series Autosampler Reference Manual 125
Repairing the Autosampler 4
3 Remove the two fixing bolts from the base of the metering
head assembly.
4 Disassemble the metering head assembly.
5 Use a small screwdriver to carefully remove the seal.
Clean the chamber with lint-free cloth. Ensure all
particular matter is removed.
6 Install the new seal. Press the seal firmly into position.
CellFrame