162 Agilent 1200 Series Autosampler Reference Manual
4 Repairing the Autosampler
8 If an Agilent ChemStation is also connected, restart the
Agilent ChemStation now as well.
Recover Instructions
1 Turn off the module.
2 Change the 8-bit Configuration Switch to Resident (see “Stay-Resident
Settings” on page 256).
3 Turn the module on.
4 Re-do steps 2 to 5 of “Using the Instant Pilot G4208A” on page 161 and
correct the type information. Enter the product number without “-R”
5 Turn the module off.
6 Change the 8-bit Configuration Switch back to default settings (see “Setting
the 8-bit Configuration Switch” on page 252).
7 Turn the module ON again. The Maintenance screen should display the
correct type for this module.
Using the Control Module G1323B
1 Connect the control module to the detector. Turn ON the detector.
2 On the control module, press System (F5), then Records (F4). Using the
up/down arrows, make sure that the detector is highlighted.
3 Press FW Update (F5), then m. This will display a box which says Update Enter
Serial#.
4 Press Enter. This will display the box labeled Serial#.
5 Letters and numbers are created using the up and down arrows. Into the
box labeled Serial#, enter the 10-character serial number for the detector.
When the 10-character serial number is entered, press Enter to highlight
the complete serial number. Then, press Done (F6).
6 Turn the detector OFF then ON again. The Records screen should display
the correct serial number for this module.
7 If an Agilent ChemStation is also connected, restart the
Agilent ChemStation now as well.