Agilent 1200 Series Autosampler Reference Manual 73
Troubleshooting and Test Functions 3
Needle Up Failed
The needle arm failed to move successfully from the seat or out of the vial to
the upper position.
The upper position of the needle arm is monitored by a position sensor on the
sampling unit flex board. The sensor detects the successful completion of the
needle movement to the upper position. If the needle fails to reach the end
point, or if the sensor fails to recognize the needle arm movement, the error
message is generated.
Probable Causes
Defective or dirty position sensor.
Defective motor.
Sticking spindle assembly.
Defective ASM board.
Suggested Actions
Exchange the sampling unit flex board.
Exchange the needle drive motor.
Exchange the spindle assembly or sampling unit assembly.
Exchange ASM board.