Agilent 1200 Series Autosampler Reference Manual 59
Troubleshooting and Test Functions 3
Shutdown
An external instrument has generated a shut-down signal on the remote line.
The autosampler continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
Leak detected in an external instrument with a remote connection to the
system.
Shut-down in an external instrument with a remote connection to the
system.
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
Fix the leak in the external instrument before restarting the autosampler.
Check external instruments for a shut-down condition.
Check the degasser for an error condition. Refer to the Reference Manual
for the Agilent 1200 Series degasser.