Agilent 1200 Series Autosampler Reference Manual 63
Troubleshooting and Test Functions 3
Leak Sensor Open
The leak sensor in the autosampler has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable Causes
• Leak sensor not connected to the ASM board.
• Defective leak sensor.
Suggested Actions
✔ Ensure the leak sensor is connected correctly.
✔ Exchange the leak sensor.