Agilent 1200 Series Autosampler Reference Manual 161
Repairing the Autosampler 4
Where: XXXXX is the 5-character product number of the module
(e.g. G1314A).
2 Turn OFF the module, then ON again. Then, restart the
Agilent ChemStation. If the serial number you have just entered is different
than the original module serial number, you will be given the opportunity to
edit the configure 1200 access screen during the restart of the Agilent
ChemStation.
3 After restart, the serial number/type you have just entered can be seen
under the Instrument menu of the main user interface screen.
Using the Instant Pilot G4208A
1 Connect the Instant Pilot to the module. Turn ON the module.
2 On the Instant Pilot’s Welcome screen, press More, then select Maintenance.
Using the up/down arrows, select the module where you have to change the
product number or serial number.
3 Press PN/SN. This will display a screen where you can enter the product
number and/or serial number.
4 Make your changes, using the information from the product label of your
module.
5 Press OK to highlight the complete command.
6 Press Done to transfer the information into the main board’s memory. Press
Cancel quit the process.
7 Turn the module OFF then ON again. The Maintenance screen should
display the correct serial number for this module.
WARNING
If you enter the wrong type, your module will not be accessible anymore. In such
a case see “Using the Instant Pilot G4208A” on page 161 or “Using the Control
Module G1323B” on page 162 for recovering.
WARNING
If you enter the wrong type, your module might not be accessible anymore with
the Agilent ChemStation and the Instant Pilot (unsupported module). In such a
case follow the “Recover Instructions” on page 162.