134 Agilent 1200 Series Autosampler Reference Manual
4 Repairing the Autosampler
Top Cover and Foam
Frequency When accessing internal parts.
Tools required If interface board installed: Flat-head screwdriver.
Parts required Foam kit G1313-68702
CAUTION
This procedure requires removal of the MIO-interface board. The board is sensitive to
electrostatic discharge. Always use the ESD strap when handling electronic boards.
1 Switch OFF the autosampler at the main power switch.
Remove the power cable.
2 If installed, remove the ECB board.