EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1200 series

Agilent Technologies 1200 series Reference Manual

Agilent Technologies 1200 series
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
Agilent 1200 Series Autosampler Reference Manual 85
Troubleshooting and Test Functions 3
User Interface
The functions for the ChemStation and Control Module (different names for
functions in the Control Module are shown in brackets) are:
Change Needle:
moves the safety flap away from the needle, and positions the needle arm for
easy access to the needle and needle seat.
Change Piston:
relieves the tension on the metering spring (draws the piston to the outer
position), enabling easy disassembly of the metering head assembly).
Park Arm (Park Gripper):
secures the gripper arm to the park position behind the sampling unit. ready
for transport or shipping of the autosampler.
Home:
moves the tray arm to its home position for better access and exchange of the
trays.
Change Gripper:
The change gripper function moves the gripper to the front of the autosampler
enabling easy access to the gripper release mechanism.

Table of Contents

Other manuals for Agilent Technologies 1200 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 series and is the answer not in the manual?

Agilent Technologies 1200 series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals