EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1200 series

Agilent Technologies 1200 series Reference Manual

Agilent Technologies 1200 series
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Agilent 1200 Series Autosampler Reference Manual 67
Troubleshooting and Test Functions 3
Fan Failed
The cooling fan in the autosampler has failed.
The hall sensor on the fan shaft is used by the ASM board to monitor the fan
speed. If the fan speed falls below 2 revolutions/second for longer than
5 seconds, the error message is generated.
Probable Causes
Fan cable disconnected.
Defective fan.
Defective ASM board.
Suggested Actions
Ensure the fan is connected correctly.
Exchange fan.
Exchange the ASM board.

Table of Contents

Other manuals for Agilent Technologies 1200 series

Questions and Answers:

Agilent Technologies 1200 series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals