62 Agilent 1200 Series Autosampler Reference Manual
3 Troubleshooting and Test Functions
Leak
A leak was detected in the autosampler.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the ASM board.
Probable Causes
• Loose fittings.
• Broken capillary.
• Leaking rotor seal or needle seat.
• Defective metering seal.
Suggested Actions
✔ Ensure all fittings are tight.
✔ Exchange defective capillaries.
✔ Exchange the rotor seal or seat capillary.
✔ Exchange the metering seal.