4 Repairing the Autosampler
Gripper Arm
Frequency Defective gripper arm
Tools required Straightened paper clip.
Parts required Gripper assembly, G1313-60010.
Preparations for this
procedure
Select “Start” in the maintenance function “ChangeGripper” (see “Change
Gripper (Change Arm)” on page 90).
Turn off the power to the autosampler.
Remove the front cover.
1 Identify the slit below the gripper motor and the gripper
arm release button.
2 Rotate the arm approximately 2.5cm (1 inch) to the left and
insert the straightened paper clip into the slit.
.
CellFrame
slit
button
CellFrame