14 Agilent 1200 Series Autosampler Reference Manual
1 Installing the Autosampler
Site Requirements
A suitable site environment is important to ensure optimal performance of the
autosampler.
Power Consideration
The autosampler power supply has wide-ranging capability. Consequently
there is no voltage selector in the rear of the autosampler. There are also no
externally accessible fuses, because automatic electronic fuses are
implemented in the power supply.
The thermostatted autosampler comprises two modules, the standard or
preparative autosampler and the thermostat (G1330B). Both modules have a
separate power supply and a power plug for the line connections. The two
modules are connected by a control cable and both are turned on by the
autosampler module.
WARNING
To disconnect the autosampler from line power, unplug the power cord. The power
supply still uses some power, even if the power switch on the front panel is turned
off.
WARNING
To disconnect the thermostatted autosampler from line power, unplug the power
cord from the autosampler and the ALS thermostat. The power supplies still use
some power, even if the power switch on the front panel is turned off. Please make
sure that it is always possible to access the power plug.
WARNING
Shock hazard or damage of your instrumentation can result if the devices are
connected to a line voltage higher than specified.