EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1200 series

Agilent Technologies 1200 series Reference Manual

Agilent Technologies 1200 series
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
Agilent 1200 Series Autosampler Reference Manual 81
Troubleshooting and Test Functions 3
Vial in Gripper
The gripper arm attempted to move with a vial still in the gripper.
During specific stages of the sampling sequence, no vial should be held by the
gripper. The autosampler checks if a sample vial is stuck in the gripper by
closing and opening the gripper fingers while monitoring the motor encoder. If
the gripper fingers are unable to close, the error message is generated.
Probable Causes
Vial still in gripper.
Suggested Actions
Remove the vial using the “Release Vial” function in the user interface.
Reinitialize the autosampler.

Table of Contents

Other manuals for Agilent Technologies 1200 series

Questions and Answers:

Agilent Technologies 1200 series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals