EasyManua.ls Logo

Agilent Technologies 1200 series - Synchronization Lost

Agilent Technologies 1200 series
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1200 Series Autosampler Reference Manual 61
Troubleshooting and Test Functions 3
Synchronization Lost
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Probable Causes
CAN cable disconnected.
Defective CAN cable.
Defective main board in another module.
Suggested Actions
Ensure all the CAN cables are connected correctly.
Switch off the system. Restart the system, and determine which module or
modules are not recognized by the system.
Ensure all CAN cables are installed correctly.

Table of Contents

Other manuals for Agilent Technologies 1200 series

Related product manuals