EasyManuals Logo

Agilent Technologies 1290 infinity User Manual

Agilent Technologies 1290 infinity
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
1290 Infinity Quaternary Pump User Manual 137
Test Functions and Calibrations
8
Pump Leak Rate Test
Troubleshooting the Pump Leak Rate Test
Secondary Leak > 3 µL/min
Leak between the OBV and automatic purge valve
Check for visible leaks on fittings and connectors
Connector not fixed / tight enough
Connector damaged
Leaky filter frit assembly
Remove the seal wash tubes from the support ring and check for leak
into the seal wash path
Main seal leaking/damaged
Piston damaged
Remove waste lines from the automatic purge valve
Damage to rotor seal and/or stator head
Outlet valve not properly assembled
Re-tighten the outlet valve
Check the position of the gold seal
Dynamic Leak > 30 µL/min or Dynamic Leak Rate Test fails
Air in the primary pump chamber
Check for air in the solvent inlet lines and the Tuning signal
Purge the lines, Prime and Condition the pump head
Abort due to over pressure
Check solvent and solvent settings
Purge and condition the pump head with water
Leak in Inlet Valve
Check for moving air bubbles in tubing directly to the Inlet Valve
Purge the lines with water to remove dirt
Knock at the valve, clean or replace it

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals