EasyManuals Logo

Agilent Technologies 7890 Series User Manual

Agilent Technologies 7890 Series
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
180 Advanced Operation Manual
5 Detectors
Setting parameters for the TCD
1 Press [Front Det] or [Back Det].
2 Set the detector temperature. Do not set higher than the
maximum temperature allowed for the column because part of
the column passes through the heated block and into the cell.
3 Verify that makeup gas type is the same as that plumbed
to your instrument (next to Makeup line in the parameter
list). Change the gas type, if necessary.
4 Set the reference gas flow rate.
5 If you are using packed columns, turn off the makeup gas
(or proceed to step 6 and enter 2 to 3 mL/min, see “TCD
carrier, reference, and makeup gas” on page 176) and
proceed to step 7
6 If you are using capillary columns:, choose a flow mode
and set the makeup gas flow or combined flow.
7 Turn on the filament. Allow about 30 minutes for thermal
stabilization. A longer period may be needed for the
highest sensitivity.
8 If necessary, turn Negative polarity [On/Yes] to invert
negative- going peaks. When a sample contains
components giving both positive- and negative- going
peaks, Negative polarity can be switched on and off during
a run as a timetable event.
Example: Packed mode (packed and large capillary columns)
Column flow is 15 to 60 mL/min. Set the reference flow to
1.5 times the sum of column flow + makeup flow.
Makeup gas is recommended with all capillary columns. It
allows the column to be inserted all the way into the
detector and withdrawn 1 mm. If makeup is not used, the
column must be no more than 3 mm above the ferrule.
Minimum makeup flow is 1 mL/min.
1/8-inch stainless steel column If column flow is 30 mL/min,
set the reference flow to 30 × 1.5 = 45 mL/min. Total
detector flow is 30 + 45 = 75 mL/min.
10 m × 0.53 mm column If column flow is 15 mL/min and
makeup flow = 2 mL/min, set the reference flow to 1.5 × (15
+ 2) = 25.5 mL/min. Total detector flow is 17 + 25.5 =
42.5 mL/min.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals