48 Advanced Operation Manual
3 Inlets
You can do column pressure and flow programming when in
the pressure pulse mode. However, the pressure pulse will
take precedence over the column pressure or flow ramp.
Split/Splitless inlet split mode minimum operating pressures
The minimum recommended inlet total flow is
20 mL/minute. When the split/splitless inlet is operated in
Split mode, there will be a minimum pressure at which the
inlet can operate. Typically, low inlet pressures may be
required for shorter, wide bore columns. The minimum
pressure is a function of carrier gas type, total inlet flow,
liner design, and possible contamination of the split vent
tube or trap.
A wide bore column requires a much lower inlet pressure
than a typical capillary column to maintain a given flow.
Setting the split ratio (total flow) too high when using a
wide bore column can create an unstable control
relationship between the pressure and flow control loops.
01 23
4
567 8
Pressure pulse
Pressure or flow program
Pressure
or
Flow

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals