Programming 1
Advanced Operation Manual 15
Clock Time Programming
Clock time programming allows certain setpoints to change
automatically at a specified time during a 24- hour day. Thus,
an event programmed to occur at 14:35 hours will occur at
2:35 in the afternoon. A running analysis or sequence has
precedence over any clock table events occurring during this
time. Such events are not executed.
Possible clock time events include:
• Valve control
• Method and sequence loading
• Starting sequences
• Initiating blank and prep runs
• Column compensation changes
• Adjustments of the detector offset
Using clock time events
The Clock Table function allows you to program events to
occur during a day based on the 24- hour clock. Clock table
events that would occur during a run or sequence are
ignored.
For example, the clock table could be used to make a blank
run before you even get to work in the morning.
Programming clock time events
1 Press [Clock Table].
2 Press [Mode/Type] to see the available clock time events.
3 Scroll to the parameter to be programmed.
4 Edit Time: and the setpoints for this event.
5 Press [Mode/Type] to add another event. Press [Status] to
terminate entries.
When the clock event is executed, a confirming message
appears.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals