184 Advanced Operation Manual
5 Detectors
• Include a letter stating the condition of abuse.
It is unlikely, even in this very unusual situation, that
radioactive material will escape the cell. However, permanent
damage to the
63
Ni plating within the cell is possible, and
therefore, the cell must be returned for exchange.
Safety precautions when handling uECDs
• Never eat, drink, or smoke when handling uECDs.
• Always wear safety glasses when working with or near
open uECDs.
• Wear protective clothing such as laboratory jackets, safety
glasses, and gloves, and follow good laboratory practices.
Wash hands thoroughly with a mild non- abrasive cleaner
after handling uECDs.
• Cap the inlet and outlet fittings when the uECD is not in
use.
• Connect the uECD exhaust vent to a fume hood or vent it
to the outside. See the latest revision of title 10, Code of
Federal Regulations, part 20, (including appendix B) or
the applicable State regulation. For other countries,
consult with the appropriate agency for equivalent
requirements.
Agilent Technologies recommends a vent line inside
diameter of 6 mm (1/4 inch) or greater. With a line of
this diameter, the length is not critical.
WARNING
Do not use solvents to clean the uECD.
WARNING
You may not open the uECD cell unless authorized to do so by your
local nuclear regulatory agency. Do not disturb the four
socket-head bolts. These hold the cell halves together. United
States customers removing or disturbing them is a violation of the
terms of the exemption and could create a safety hazard.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals