Columns and Oven 4
Advanced Operation Manual 165
Nickel Catalyst Tube
About the nickel catalyst tube
The Nickel Catalyst Tube accessory, G2747A, is used for
trace analysis of CO and CO
2
with a flame ionization
detector. The gas sample is separated on the column and
passed over a heated catalyst in the presence of hydrogen,
which converts the CO and CO
2
peaks to CH
4
.
Nickel catalyst gas flows
For a standard FID installation:
Carrier gas
Sample Hydrogen Air
Gas sample valve
Column Nickel catalyst
FID
Tabl e 40 Gas flows for a standard FID
Gas Flow rate, mL/min
Carrier (helium) 30
FID hydrogen 30 (see Caution)
FID air 400