164 Advanced Operation Manual
4 Columns and Oven
The Run Table contains commands to perform these actions:
After the last peak of interest appears, switch the valve
to Position 2. Higher boiling peaks are discarded through
the inlet vent.
At the same time, turn data acquisition off.
At the end of the backflush period (determined
experimentally), switch the valve to Position 1 and turn
data acquisition on. The system is now ready for the next
run.