Inlets 3
Advanced Operation Manual 39
To use gas saver
1 Press [Front Inlet] or [Back Inlet].
2 Turn gas saver On.
3 Set Gas saver flow. It must be at least 15 mL/min greater
than the column flow.
4 If in split mode, set Saver time after injection time. In all
other modes, set after Purge time.