Valves 6
Advanced Operation Manual 219
Valve Types
The possible valve types are:
Sampling A two- position (load and inject) valve. In load
position, an external sample stream flows through an
attached (gas sampling) or internal (liquid sampling) loop
and out to waste. In inject position, the filled sampling loop
is inserted into the carrier gas stream. When the valve
switches from Load to Inject, it starts a run if one is not
already in progress. See the example on page 222.
Switching A two-position valve with four, six, or more
ports. These are general- purpose valves used for such tasks
as column selection, column isolation, and many others. For
an example of valve control, see page 223.
Multiposition Also called a stream selection valve. It is
usually used to select one from a number of gas streams and
feed it to a sampling valve for analysis. It has a special
actuator that advances the valve one position each time it is
activated, or it may be motor driven. An example that
combines a stream selection valve with a gas sampling valve
is on page 223.
Other Could be anything.
Not installed Self- explanatory.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals