Inlets 3
Advanced Operation Manual 95
Setting COC inlet parameters
Track oven mode
1 Press [Front Inlet] or [Back Inlet].
2 Press [Mode/Type] and select Track oven.
There is no setpoint for Track oven mode.
Ramped temperature mode
1 Press [Front Inlet] or [Back Inlet].
2 Press [Mode/Type] and select Ramped temp.
3 Enter a value for Te mp . This is the starting temperature.
4 Enter an Init time. This is the length of time the inlet will
stay at the starting temperature after a run has begun.
5 Enter Rate 1. This is the rate at which the inlet will be
heated or cooled. A Rate of 0 halts further programming.
6 Enter Final temp 1. This is the inlet temperature at the end
of the first ramp.
7 Enter Final time 1. This is the number of minutes the inlet
holds Final temp 1.
8 To enter a second (or third) ramp, scroll to the
appropriate Rate line and repeat steps 5 through 7.