Auxiliary Devices 8
Advanced Operation Manual 243
Setting parameters for auxiliary pressure control
1 Press [Aux EPC #] and scroll to the channel you wish to
control. Press [Enter].
2 Scroll to Initial pressure. Enter a value and press [Enter].
3 If desired, enter a pressure program using the time and
rate functions.
4 Press [On/Yes] to apply Initial pressure and start the
program.
About Aux Thermal Zone Control
There can be up to 3 aux thermal controlled zones.
These zones are 3- ramp programmable.
Run time events may be used to schedule specific
temperatures during the run.
Setting parameters for the aux thermal zone control
1 Press [Aux Temp #] and scroll to the zone you wish to
control. Press [Enter].
2 Scroll to Te mp era tu re. Enter a value and press [Enter].
3 Scroll to Initial time. Enter a value and press [Enter].
4 Scroll to Rate 1. Enter 0 to end the program here or a
positive value to create a temperature program.
About Cryo Trap Control
When configured as a cryo trap, an aux thermal zone has
the following parameters:
Tempera ture Displays the current trap temperature and
setpoint. Press [On/Off] to turn the zone on or off.
If On, the cryo trap temperature will remain 10 °C above
the current oven temperature, following the oven program.
If Off, the GC does not control the trap temperature.
Cold Trap Temperature Enter the target temperature for cryo
trap operation and press [Enter]. Enable this temperature
using run time On and Off events for the aux thermal zone.