200 Advanced Operation Manual
5 Detectors
About the FPD
+
The sample burns in a hydrogen- rich flame, where some
species are reduced and excited. The gas flow moves the
excited species to a cooler emission zone above the flame
where they decay and emit light. A narrow bandpass filter
selects light unique to one species, while a shield prevents
intense carbon emission from reaching the photomultiplier
tube (PMT).
The light strikes a photosensitive surface in the PMT where
a light photon knocks loose an electron. The electron is
amplified inside the PMT for an overall gain of up to a
million.
The current from the PMT is amplified and digitized by the
FPD
+
electronics board. The signal is available either as a
digital signal on the communications output or as a voltage
signal on the analog output.
The FPD
+
should not be stored at temperatures above 50 °C,
based on the original manufacturer’s specifications for the
PMT.
Valve
Frit
Frit
Valve
PS = Pressure Sensor
Column
PS
PS
PS
Valve
Frit
Restrictor Restrictor
Restrictor
EPC Module
H
2
Air
Wavelength filter
Window
Vent
Makeup
Emission Zone
Transfer line
Restrictor

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals