Valves 6
Advanced Operation Manual 221
Controlling a Valve
From the keyboard
Valves (except multiposition valves) have two positions
controlled by the [On] and [Off] keys. The keyboard
commands for two- position valves are:
[Valve #] <scroll to the valve> [On] Rotates valve to one
stop
and
[Valve #] <scroll to the valve> [Off] Rotates valve to the
other stop
From the run or clock time tables
The Valve On and Valve Off commands can be run time or clock
time programmed. See “Run Time Programming” on page 12
and “Clock Time Programming” on page 15.
If a valve is rotated by a run time program, it is not
automatically returned to its initial position at the end of
the run. You must program this reset operation yourself.
Simple valve: column selection
This is the plumbing for a single valve, configured as a
switching valve, that selects one of two columns for analysis.
It has no configuration parameters.
The column is selected by pressing:
[Valve #] <scroll to valve #> [On] (for the front column) or
[Off] (for the back column).
Front column
Back column
To d ete cto r
From inlet or
sampling valve
ON
OFF

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals