GC Output Signals 7
Advanced Operation Manual 233
Data rates with Agilent data systems
The GC can process data at various data rates, each
corresponding to a minimum peak width. The table shows
the effect of data rate selection.
Baseline upset due to valve switch
Pause signal here
Resume signal here
Tabl e 58 EZChrom/ChemStation data processing
Data rate, Hz Minimum peak
width, minutes
Relative
noise
Detector Column type
500 0.0001 5 FID Narrow-bore,
0.05 mm
200 0.001 3.1 FID Narrow-bore,
0.05 mm
100 0.002 2.2 FID/FPD/NPD
only
Capillary
50 0.004 1.6
20 0.01 1