232 Advanced Operation Manual
7 GC Output Signals
Digital Signals
The GC outputs digital signals only to an Agilent data
system. The following discussions describe features that
impact the data sent to data systems, not the analog data
available to integrators. Access these features from the
data system. These features are not accessible from the
GC keypad.
Digital zero
Available only from an Agilent data system.
Digital signal outputs respond to a zero command by
subtracting the signal level at the time of the command from
all future values.
Signal Freeze and Resume
Available only from an Agilent data system.
Some run time operations, such as changing signal
assignments or switching a valve, can cause baseline upsets.
Other factors can cause baseline upsets also. The GC can
compensate for this by pausing (freezing) the signal at a
particular value, using that signal value for a specified
duration, and then resuming normal signal output.
Consider a system that uses a switching valve. When the
valve switches, an anomaly occurs in the baseline. By
freezing and resuming the signal, the anomaly can be
removed so that the peak identification and integration
software operates more smoothly.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals