EasyManuals Logo

Agilent Technologies 7890 Series User Manual

Agilent Technologies 7890 Series
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #109 background imageLoading...
Page #109 background image
Inlets 3
Advanced Operation Manual 109
8
Press [Prep Run] (see “Pre Run and Prep Run” on page 40)
before manually injecting a sample. This is automatic if
an Agilent sampler is used.
PTV inlet solvent vent mode
This mode is typically used for large volume injections. For
single injection use a 50 or 100 µL syringe with variable
plunger speed—slowly, 5 to 30 seconds.
The sample is injected into a cold inlet. If conditions are
properly chosen and the sample is suitable, analytes deposit
in the inlet liner while the solvent evaporates and is swept
out. Large or multiple injections can be used to concentrate
sample in the inlet before transferring to the column for
analysis.
The figure shows the flows with the septum head. Flows
with the septumless head are the same except that the
septum purge flow bypasses the head.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals