GC Output Signals 7
Advanced Operation Manual 231
Selecting fast peaks (analog output)
1 Press [Config][Analog 1] or [Config][Analog 2].
2 Scroll to Fast peaks and press [On].
Agilent does not recommend using Fast peaks with a thermal
conductivity detector. Since the gas streams switch at 5 Hz,
the gain in peak width is offset by increased noise.