16
Installing the Autosampler
Site Requirements
Environment
Your autosampler will work within specifications at ambient temperatures
and relative humidity as described in Table 1.
CAUTION Do not store, ship or use your autosampler under conditions where
temperature fluctuations may cause condensation within the autosampler.
Condensation will damage the system electronics. If your autosampler was
shipped in cold weather, leave it in its box, and allow it to warm up slowly
to room temperature to avoid condensation.
WARNING Using the autosampler at environmental temperatures higher than 50 °C
(122 °F) may cause the rear panel to become hot.
Table 1 Physical Specifications - Autosampler (G1313A / G1329A / G1389A / G2260A)
7\SH 6SHFLILFDWLRQ &RPPHQWV
Weight 14.2 kg (31.3 lbs)
Dimensions
(height × width × depth)
200 × 345 × 435 mm
(8 × 13.5 × 17 inches)
Line voltage
100 – 120 or 220 – 240 VAC,
± 10 %
Wide-ranging capability
Line frequency
50 or 60 Hz,
± 5%
Power consumption (apparent power)
Power consumption (active power)
300 VA
200 W
Maximum
Maximum
Ambient operating temperature
4 – 55
°C (41 – 131 °F)
see WARNING on page 16
Ambient non-operating temperature
-40 to 70
°C (-4 to 158 °F)
Humidity
<95%, at 25– 40
°C (77 – 104 °F)
Non-condensing
Operating Altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storing the autosampler
Safety standards: IEC, CSA, UL Installation Category II, Pollution Degree 2