4-22 Service Guide E8361-90001
Troubleshooting PNA Series Microwave Network Analyzers
Rear Panel Troubleshooting E8361A
Checking the SERIAL (RS-232), PARALLEL (1284-C), or VGA Port
To verify the proper operation of the SERIAL, PARALLEL, or VGA port:
Connect a known good serial, parallel, or VGA peripheral device.
Wait 15 seconds for the analyzer to verify the device connection, and then check the
operation of the peripheral device.
If the peripheral device performs correctly, the port is functioning properly.
If the peripheral device does not function properly, the A15 CPU board is faulty. Refer to
“Removing and Replacing the A15 CPU Board” on page 7-28.