Service Guide E8361-90001 B-3
PNA Series Microwave Network Analyzers Option Enable Utility
E8361A Accessing the Option Enable Utility
Accessing the Option Enable Utility
To start the option enable application:
•On the
System menu, point to Service, and then click Option Enable.
• The dialog box illustrated in Figure B-1 is displayed.
NOTE Some applications require a keyword that is provided by Agilent. If you do not
have the required keyword, contact Agilent for assistance. Refer to
“Contacting Agilent” on page 2-9.
Figure B-1 Option Enable Dialog Box
Installing or Removing Options
There are two types of options:
• Hardware option: This option involves some physical change to the analyzer. The option
enable utility allows you to load the software needed for the changes.
• Software option: This option involves loading software to initialize your new option.
The option enable utility is part of this process.
Use the
Select Desired Option box to select an option to install or remove from your
analyzer. If the option is not available in the list box, select Enter Unlisted Option. A dialog
box will allow you to enter the option number. Enter the option and follow the instructions
on the display.