Horizontal Δt and 1/Δt Accuracy Verification Test 
This test verifies the horizontal Δt and 1/ Δt accuracy. In this test, you 
will use the oscilloscope to measure the output of a time mark generator. 
Test limits: ±3% reading ±0.4% screen
Refer to Table 7- 1 for the equipment required for this horizontal Δt and 
1/  Δt accuracy verification test.
1 Connect the calibrator to the instrument’s channel 1 using a 50 
Ω 
feedthrough termination. 
2 Set up the signal source.
• Select Marker on the calibrator
• Set the calibrator for 100 
ms markers. 
• Set the spike shape on the calibrator.
3 Set up the oscilloscope. 
• Press the Autoscale key. 
• Set the time base to 20 
ms/div. 
• Select Scope mode and press F3 to access Delay menu.
• Set the Time Reference to Left at F2. 
• Adjust the trigger level to obtain a stable display. 
4 Select Measure mode, then press F1 to enable auto measurement 
selection for Measure 1. Turn rotary switch to select Frequency 
measurement from the list of auto measurements. Set Measure 2 for 
Period measurement. Measure the following: 
• Frequency of 10 kHz — Test limits are 9.63 kHz to 10.40 kHz. 
• Period of 100 
ms — Test limits are 96.20 ms to 103.8 ms.
5 Change the calibrator to 1 
µs markers and time base to 250 ns/div. 
Adjust the trigger level to obtain a stable display. 
6 Measure the following:
• Frequency of 1 MHz — Test limits are 962 kHz to 1.04 MHz. 
• Period of 1 
ms — Test limits are 0.96 ms to 1.04 ms.