Sequoia Series    185 
8.2  IEC 1000-4-11 (-411) 
The following are the remote commands set for IEC1000-4-11. The ABORt command will terminate a 
test in progress at any time. Test completion can be checked using the *OPC? query if the *OPC 1 
command is issued immediately after an IEC1000-4-11 command. Upon completion of the test run 
*OPC? will return a 1.  This may be relevant for ATE applications in view of long test times for some 
IEC 1000-4-11 test sequences. 
Details on the actual tests performed as a result of each of these commands may be found in the IEC 
1000-4-11 option section of the User Manual which is supplied with the power source. 
IEC411: 
  [:NOMinal] 
    :VOLTage 
  :DIPS 
    :RUN 
      [:ALL] 
      :SINGle 
    :VOLTage 
      [:PERCent] 
    :CYCLes 
    :ANGLe 
  :STATe 
  :VARiants 
    :RUN 
      [:ALL] 
      :SINGle 
    :HOLD 
      :TIME 
      :VOLTage 
        [:PERCent] 
    :FALL 
      [:TIME] 
    :RISE 
      [:TIME] 
 
IEC411[:NOMinal]:VOLTage 
This command set the nominal voltage for the test 
Command Syntax  IEC411[:NOMinal]:VOLTage <NRf> 
Parameters  <voltage>  
Examples  IEC411:VOLT  230 
Query Syntax  IEC411:VOLT? 
Returned Parameters  <NR2>