100 Sequoia Series
Command Syntax [SOURce:]LIST:CURRent:SLEW <NRf+>,<NRf+>
Parameters 0.1 to 1E9 | MAX
Unit V/S (volts per second)
Example LIST:CURR:SLEW 10, 1E2, MAX
Query Syntax [SOURce:]LIST:CURRent:SLEW?
Returned Parameters <NR2>
Related Commands LIST:CURR:SLEW:POIN? LIST:COUN
LIST:DWEL LIST:STEP
LIST:CURRent:SLEW:POINts?
Phase Selectable
This query returns the number of points specified in LIST:CURRent:SLEW. Note that it returns only
the total number of points, not the point values.
Query Syntax [SOURce:]LIST:CURRent:SLEW:POINts?
Returned Parameters <NR1>
Example LIST:CURR:SLEW:POIN?
Related Commands LIST:CURR:SLEW
4.22 Source Subsystem - Mode
The MODE command allows switching between the different output modes available on the power
source. When switching modes, the output must be turned OFF. After a mode command, the output
voltage needs to be programmed to the desired setting.
Subsystem Syntax
[SOURce:]
MODE AC | DC | ACDC Sets the output mode
MODE
The mode command switches the output voltage between the available output modes.
Command Syntax [SOURce:]MODE
Parameters AC | DC | ACDC
Example MODE AC
Query Syntax [SOURce:]MODE?
Returned Parameters <CRD>
Related Commands PONS:VOLT:MODE